Euterpe features detailed

Euterpe Single-Use-Pump's (SUP) all consist of:

The Euterpe Air-Operated-Diaphragm-Positive-Displacement (AODPD) Single-Use-Pump (SUP) cell’s consist of a housing integrating a valve body, a dome, the diaphragm, and the fluid conveying hoses – all supplied pre-sterilized in dual film bags.

Stand-alone product, model names

Euterpe-30 

Euterpe-100 

Euterpe-300 

Euterpe-500 

Layout, number of diaphragms

single

single

single

single

Diaphragm diameter, mm

 60  80  125  150

Diaphragm stroke, mm

 25  39  46  54

Stroke Volume, SV, mL / 100 % stroke

 30

 100

 300

 500

Cardiac Output, CO range, mL / min

0 - 750

0 - 1,000

0 - 3,000

0 - 5,000

Beats-per-Minute, BpM

0 - 30

0 - 20

0 - 15

0 - 14

Typical hose connection, ID/OD mm

4/8

8/12

12/18

12/18

SUP cell weight, grams 100 200 300 400

The basics in operation is copied from the human heart functionality:

  • Cardiac Output (CO) = total pumped volume, SUP capacity, ml/min (CO = BpM x SV)
  • Beats-per-Minute (BpM) = ranging 0 - 30 BpM, SUP diameter/stroke/volume depending
  • Stroke Volume (SV) = programmable from 10 - 100% of SUP cell specification
Red Laser - foot - Euterpe-100.JPGD3C_2267.JPGEuterpe-100 kit.JPGfoot and Euterpe-100.JPGLachesis-2 control unit.JPG

Physical features

  • No expensive housing materials for traditional steam cleaning requirements
  • No mechanical axle seals
  • No mechanical rotating parts such as electrical motor, coupling or gears
  • No moving ball valves, only passive cross-slit silicone valves
  • All simple single-use plastic parts – contamination eliminated
  • Low weight and compact design gives OEM manufactures new freedom to design equipment
  • Measures accurately pumped volume, a true Positive Displacement pump – no in-line flow-sensor is needed
  • All pumps offer infinite turn-down, turn-up scale of 1:5.000

Operating features

  • Volumetric operation precision better than 1% - without any extra flow sensor
  • Volume per stroke depending on pump size - no minimum volume exist
  • Zero spallation / shedding of membrane materials - shear never introduced
  • Self-priming from 2 meter suction height
  • Standard fluid pressure 1 bar - higher pressure like 3-5 Bar require customized body
  • No need for calibration - build in artificial inteligence in Drive Unit
  • Pump any fluid - any mixture of liquid and gases
  • Dry running for extended periods
  • Self-emptying, drain on inlet and purge on outlet as the pump will convey gas as well
  • Single-use, disposable wetted parts eliminate cleaning issues – just un-pack and exchange without any tools needed
  • Very gently operation with no rotating or rubbing parts - no shear introduced between used polymeric materials parts creating nano size polluting particles
  • Working pressure as much as 5 Bar on special models makes the SUP suitable for connection with filtration devices
  • Two or four pumps able to operate fully controlled and in parallel connected to Atropos-2
  • Two pumps able to operate fully controlled in parallel when connected to Atropos-1

Programming features (on Atropos)

  • Select the continuous volume you want to pump – (well, according to pump size)
  • Select the total volume you want to pump
  • Select the time you want the pump to operate
  • Real-time information about pumping, pumped volume and velocity

Maintenance features

  • No tools needed replacing the SUP cell - takes only one minute !
  • No lubrication needed
  • No need to strip down pump for cleaning
  • Longevity - tested to last 1 million cycles at 75% CO

Installation features

  • Single-Use-Pump cell may easily be part of an all sterile process package
  • No tooling needed for Single-Use-Pump cell replacement - one minute operation

Data collection features of Atropos Drive Unit

  • On-line operation pressure measured by build-in pressure sensor
  • On-line Mass-Flow-Control / metering, the pump measures the volume pumped with better than 1% accuracy - without external flow sensors!
  • Data acquisition, data logging

Communication features of Atropos drive unit

  • Single-Use-Pump cell's need connection to Atropos Drive Unit (requires only pressurized air, vacuum and 24 VDC electrical power)
  • TCP/IP bus connection or Wi-Fi connection
  • All Euterpe kids are stand-alone and share the same family of Atropos Drive Unit's
  • Atropos software up-gradable
  • Atropos units facilitate a webhost, which allow the user to access Atropos via Wi-Fi as well as via IP/TCP from many devices via Ethernet, USB connection (for data export), OPC communication
  • Access Atropos via build-in touch sensitive display or connect via Wi-Fi a PAD, Smartphone or PC

Not designed for:

  • Viscosity higher than 10,000 centiStoke
  • Very abrasive materials for a long time